Home > About Us > Bettersize Newsroom > News > Top Particle Analysis Systems to Streamline Regulated Pharmaceutical Workflows

Top Particle Analysis Systems to Streamline Regulated Pharmaceutical Workflows

2026-11-27News

Bettersize Science Blogs
PHARMACEUTICAL PARTICLE ANALYSIS

Top Particle Analysis Systems to Streamline Regulated Pharmaceutical Workflows

Choosing pharmaceutical particle analysis systems: from nanoformulation screening to particle size QC and data review

When screening lipid nanoparticle (LNP) formulations, you need to understand how particle size and dispersion state respond to formulation changes. When evaluating a micronized active pharmaceutical ingredient (API), the questions shift to the overall size distribution, changes at its coarse end, and particle morphology. In quality control (QC), these results must also be accompanied by sample preparation details, method versions, and review records.

ANALYTICAL TECHNIQUE
Dynamic light scattering · Laser diffraction · Dynamic image analysis

The right pharmaceutical particle analysis system brings measurement data, routine operation, and data review together around a defined quality question. Bettersize pharmaceutical solutions cover R&D, process development, and QC, helping laboratories choose characterization methods for their samples and tasks.

Key takeaway: The BeNano 180 Zeta Max supports particle size and zeta potential studies of nanoscale dispersions. The Bettersizer 2600 Plus uses interchangeable wet and dry laser diffraction configurations for API and excipient powders and suspensions, with optional PIC-1 imaging to extend particle characterization. Combining suitable methods, repeatable procedures, and traceable electronic records helps carry R&D methods into routine testing.

Two priorities to evaluate before selecting a system

1. Match the measurement principle to the quality question

Dynamic light scattering (DLS) analyzes fluctuations in scattered light intensity caused by Brownian motion. It provides diffusion information, which is used with temperature and medium viscosity to calculate hydrodynamic diameter. DLS is suited to nanoscale dispersions, including proteins and liposomes, and to tracking how size and distribution respond to formulation or environmental conditions.

Laser diffraction (LD) uses scattered light from a particle population and an inversion model based on Mie theory or the Fraunhofer approximation to calculate a particle size distribution. It supports comparisons of APIs, excipients, and suspensions. Dynamic image analysis (DIA) records two-dimensional particle projections, adding shape information and individual particle images.

Use DLS to track nanoscale dispersion behavior, LD to compare the overall particle population, and DIA to examine particle dimensions and morphology. They report hydrodynamic diameter, optically equivalent spherical diameter, and projected dimensions, respectively. Interpret each result according to its measurement principle and use the methods as complementary sources of information.

2. Establish sample preparation conditions that support meaningful comparisons

Dispersion conditions determine the state in which a powder reaches the measurement zone. For wet methods, evaluate the medium, circulation, ultrasonication, and measurement time, including the effects of dissolution, settling, or particle breakage. For dry methods, optimize dispersion pressure and feeding conditions to balance deagglomeration with particle integrity.

DLS requires control of concentration, temperature, viscosity inputs, and sample handling. A dilution medium can change the particles' chemical environment. Use comparable conditions when evaluating formulations and document adjustments made for individual samples.

Establish these conditions with representative samples, then incorporate them into a reproducible, transferable method. This creates a common basis for later batch and formulation comparisons.

BeNano 180 Zeta Max: informing nanoformulation screening

The BeNano 180 Zeta Max provides nanoscale particle sizing and zeta potential characterization on one platform for liposome, LNP, nanosuspension, and protein formulation research. Researchers can combine size distribution and charge-related information to compare formulations, track responses to changing conditions, and select samples for further investigation.

BeNano 180 Zeta Max

Track formulation changes through particle size and distribution

The BeNano 180 Zeta Max has a nominal particle size range of 0.3 nm–15 μm, with suitability depending on the sample and accessories. Backscattering DLS with an adjustable detection position reduces the effects of multiple scattering in concentrated samples, supporting size analysis under appropriate sample conditions.

When comparing formulations, review the Z-average hydrodynamic diameter, polydispersity index (PDI), and measurement quality information together. Size describes the overall dimensional trend, while PDI describes the breadth of the distribution characterized by DLS. Together, they help identify changes worth tracking.

For early formulations with limited sample availability, a capillary sizing cell that accepts 3–5 μL supports small-volume particle size measurements, reducing the sample required for this step. Prepare samples for zeta potential and other measurements according to the requirements of their respective cells.

Add zeta potential to the assessment of colloidal behavior

The BeNano 180 Zeta Max measures zeta potential using electrophoretic light scattering and applies phase analysis light scattering (PALS) to analyze low electrophoretic mobility signals. Zeta potential characterizes the electrical potential associated with the particle's slipping plane and supports comparisons of charge-related interactions as conditions change.

When studying the effects of pH or ionic conditions, compare size and zeta potential together: size tracks dimensions and distribution, while zeta potential adds charge-related information. For systems stabilized with polymers or surfactants, include mechanisms such as steric stabilization when establishing formulation-specific evaluation criteria.

The optional BAT-1 autotitrator organizes pH adjustment and measurement into a continuous sequence to obtain size or zeta potential trends as a function of pH. This helps maintain a consistent procedure when repeatedly comparing different pH conditions.

Build comparable time series for stability studies

Tracking particle size over time and temperature helps researchers compare changes in the physical state of formulations. Expose samples to consistent storage or temperature conditions, then compare size and distribution using the same method to turn individual measurements into comparable trend data.

These trends guide the selection of formulations for further study and inform stability investigations. Shelf-life and quality conclusions draw on the project's defined stability program and other quality attributes.

Bettersizer 2600 Plus: flexible wet and dry sizing with optional imaging

The Bettersizer 2600 Plus uses modular wet and dry dispersion to address the particle sizing needs of APIs, excipients, and suspensions. Dry powders can use a suitable gas-based dispersion method, while suspensions can use wet recirculation. When particle morphology or coarse particles require further examination, the optional PIC-1 dynamic imaging module combines LD and DIA in a wet workflow. This brings routine size comparisons, morphology investigations, and extended characterization together on one platform.

Bettersizer 2600 Plus

Choose the testing route to match the sample state

The Bettersizer 2600 Plus has nominal LD ranges of 0.02–2,600 μm for wet dispersion and 0.1–2,600 μm for dry dispersion. The optional PIC-1 dynamic imaging range is 2–3,500 μm. Combining PIC-1 with LD in a wet workflow extends the platform's particle characterization upper limit to 3,500 μm and adds particle dimensions, shape information, and images. Applicable ranges depend on the sample and its preparation.

For free-flowing or moderately cohesive dry APIs and excipients suited to compressed-air dispersion, the 2600 Plus supports direct sizing without liquid pre-wetting, reducing medium-preparation and liquid-waste handling steps and avoiding liquid-induced dissolution or swelling. Samples requiring liquid dispersion can use wet units for water-based media or compatible organic solvents. Choose the route according to solubility, dispersion behavior, and the measurement objective.

LD reports a volume-weighted distribution (D10, D50, and D90). If D50 remains close to its historical value while the coarse end changes, PIC-1 images can help reveal agglomerates or unusual morphology for investigation alongside dispersion conditions and process records. Use suitable sampling and particle counts when evaluating rare particles so that image observations are representative.

Address regulated workflow requirements through measurement, dispersion, and software

Measurement principle: The 2600 Plus uses LD to support pharmaceutical powder and suspension methods developed with reference to ISO 13320, USP <429>, and Ph. Eur. 2.9.31. The instrument collects light scattered at different angles and uses Mie or Fraunhofer models to obtain a volume-based size distribution. The laboratory then determines optical inputs, evaluates repeatability, and establishes method suitability for its samples.

Dispersion method: Wet and dry dispersion are both dispersion routes used in LD. The 2600 Plus lets laboratories select gas-based or liquid recirculation dispersion according to sample properties and incorporate feeding, dispersion intensity, and measurement conditions into the method. Within the pharmacopoeial framework, the central question is whether the selected route provides representative, reproducible size distributions for the intended use.

Software and electronic records: For workflows involving applicable electronic records, the 2600 Plus can be configured with software that supports compliance with FDA 21 CFR Part 11. User permissions, audit trails, electronic signatures, and data review can form part of a controlled workflow, supported by installation and operational qualification (IQ/OQ) documentation. Method standards, electronic record controls, and laboratory validation each have a defined role in implementing regulated testing.

Carry methods from R&D into production QC on one platform

During R&D, select a dispersion unit based on sample state and available sample quantity, then establish baseline size distributions for APIs, excipients, or suspensions. Add PIC-1 wet imaging when morphology needs investigation. During process development and scale-up, use controlled methods to compare distributions and coarse-end changes before and after processes such as micronization, providing measurement data to inform adjustments.

For incoming material inspection and production batch QC, incorporate the established parameters into standard operating procedures (SOPs). With suitable automated dispersion units that support SOPs, applicable alignment, dispersion, measurement, reporting, and cleaning steps can be automated to reduce repeated setup and handling. Transferring sample preparation, measurement conditions, and reporting conventions together helps R&D, process, and QC teams discuss results on a common methodological basis.

How do BeNano and Bettersizer 2600 Plus fit different pharmaceutical tasks?

Define the decision the result needs to support, then select the platform. The comparison below provides a starting point for representative-sample evaluation.

Evaluation area

BeNano 180 Zeta Max

Bettersizer 2600 Plus

Main tasks

Nanoformulation screening; dispersion state and charge-related behavior

Wet/dry sizing of API and excipient powders and suspensions; morphology investigations as needed

Measurement principles

DLS; PALS-based electrophoretic light scattering

LD; optional PIC-1 wet dynamic image analysis

Results

Hydrodynamic diameter, PDI, zeta potential

Volume-based PSD and D10/D50/D90; optional PIC-1 particle images and shape information

Nominal size ranges

0.3 nm–15 μm; suitability depends on sample and accessories

Wet LD: 0.02–2,600 μm

Dry LD: 0.1–2,600 μm

Optional DIA: 2–3,500 μm

Key sample conditions

Concentration, temperature, viscosity, sample handling

Dry: gas pressure and feeding; wet: medium and dispersion conditions; imaging: representative sampling

Configuration options

BAT-1 autotitrator for pH-dependent studies

Wet/dry units selected for the sample; optional PIC-1 wet imaging module

Teams working on both nanomedicine research and API or excipient QC can establish separate methods for the two systems and connect results through consistent sample identifiers and quality procedures. BeNano tracks nanoscale size and charge-related changes. The 2600 Plus provides wet and dry sizing for powders and suspensions, with optional PIC-1 wet imaging when needed.

Building a particle analysis workflow from R&D to routine QC

Step 1: Define the measurement objective

Specify the sample state, target parameters, and intended use of the result. For example, micronization studies focus on the overall distribution and coarse-end changes, while nanoformulation screening focuses on size, PDI, and zeta potential trends. Recording these objectives in the test plan keeps method optimization and configuration choices focused on the same question.

Step 2: Develop the method with representative samples

Include normal batches and variations relevant to the project. Examine the effects of dispersion, concentration or obscuration, temperature, and analysis settings. For dry methods on the 2600 Plus, evaluate dispersion pressure and feeding conditions. For wet imaging with the optional PIC-1, also evaluate image thresholds, particle counts, and sampling representativeness.

Use these studies to define routine operating conditions, system suitability checks, and result evaluation criteria. When introducing a new instrument or method, compare results for the intended use and document systematic differences and any necessary adjustments.

Step 3: Incorporate established conditions into controlled procedures

Include sample preparation, method versions, replicate measurement plans, cleaning, and review requirements in SOPs. During transfer from R&D to QC, have the receiving team test representative samples under the established conditions to confirm that the method can be implemented by routine personnel and procedures.

Within Bettersize pharmaceutical laboratory configurations, the BeNano 180 Zeta Max and Bettersizer 2600 Plus can be configured with software that supports compliance with 21 CFR Part 11, along with SOP workflows and installation qualification/operational qualification (IQ/OQ) documentation. Map the delivered permissions, audit trails, electronic signatures, method controls, and data review functions to laboratory procedures. Plan instrument qualification and method validation separately to define the respective responsibilities of the supplier and the laboratory.

Step 4: Keep data review aligned with measurement

Maintain links between sample identifiers, preparation conditions, instrument and software versions, raw data, and final reports. When reviewing an unexpected result, first examine measurement quality and process records, then determine investigation or retesting steps under the established procedure. This gives subsequent actions a traceable basis.

For example, if a micronized batch has a D50 close to its historical value but changes in D90 and the coarse end, particle images from the 2600 Plus wet PIC-1 configuration can help investigate agglomeration or morphology differences alongside dispersion studies and process records. If the routine method is dry, wet imaging can serve as a complementary investigation, with preparation and dispersion conditions documented for both methods. This illustrative sequence gives each stage a clear purpose: identify the change, add visual information, and investigate the conditions.

Use representative samples to find the right configuration

A streamlined pharmaceutical particle analysis workflow gives every measurement a clear purpose. The BeNano 180 Zeta Max brings nanoscale size and zeta potential information to formulation comparisons. The Bettersizer 2600 Plus offers wet and dry sizing for APIs, excipients, and suspensions, with optional PIC-1 wet imaging to add morphology and coarse-particle information. SOPs, appropriate automation, and electronic record management connect operation with review, helping laboratories organize work from R&D through production QC around their quality questions.

Start with your samples: use the Bettersize free sample testing service to submit representative sample information, your current method, and the problem you most want to solve. Discuss the measurement route and configuration with the application team. For nanoformulation screening, explore the BeNano 180 Zeta Max. For dry APIs and excipients, suspensions, or combined size and shape evaluation, explore the Bettersizer 2600 Plus and discuss wet/dry dispersion and optional PIC-1 requirements.

Bettersize free sample testing
FREE SAMPLE TESTING
Test Your Samples with Bettersize

Discuss your samples and measurement goals with our team.

Request a Free Test →

Frequently asked questions

1. Should a pharmaceutical laboratory choose BeNano or Bettersizer 2600 Plus?

Start by evaluating the BeNano 180 Zeta Max for size and zeta potential studies of nanosuspensions, liposomes, LNPs, and protein formulations. For dry APIs and excipients or micrometer-scale suspensions, focus on the Bettersizer 2600 Plus. Dry powders suitable for gas-based dispersion can be tested without liquid wetting or suspension preparation. If particle images are needed, add the optional PIC-1 for wet imaging or combined wet LD+DIA analysis.

2. Why can DLS and laser diffraction report different particle sizes?

They use different measurement principles and size definitions. DLS reports a hydrodynamic diameter related to diffusion behavior. LD uses scattered light and an inversion model to obtain a distribution of optically equivalent spherical diameters. Sample state and result weighting also affect comparisons, so interpret each result according to its method.

3. Why examine particle images when analyzing API particle size?

Particle images add morphology information to the overall size distribution. With the optional PIC-1, the Bettersizer 2600 Plus can examine captured particles in a wet workflow, investigate agglomerates or unusual morphology, and extend the platform's particle characterization upper limit to 3,500 μm, providing clues for coarse-end investigations.

4. How can size and zeta potential be used together in formulation research?

Size tracks changes in particle dimensions and distribution, while zeta potential adds information about charge-related interactions. Comparing both across the same formulations, pH values, or medium conditions helps guide further research. Establish evaluation criteria around the formulation's stabilization mechanisms and other quality attributes.

FURTHER READING  /  PSA for Pharma

Want to explore
our pharmaceutical solutions?

Bettersize’s comprehensive portfolio addresses the specific technical hurdles of each phase—accelerating stability assessments in R&D, optimizing morphological parameters during scale-up, and ensuring rheological consistency in commercial tableting. 